Delivery of Optical spectroradiometer for IR band under the following conditions with these required technical parameters: - The optical input to spectroradiometer must be diffuser with cosine correction - The spectroradiometer must measure Spectral irradiance (W/m2*nm), other measurements are an advantage e.g. irradiance (W/m2) - Measurement range of IR spectroradiometer meter must be less than 1000 nm and more than 2000 nm - Pixel resolution must be less than 2.5nm/pixel - Optical bandwidth must be less than 10 nm - Analog to digital converter of spectroradiometer must be more than or equal to 16 bit - Operation temperature range must be more than 22°C ± 5 °C - The spectroradiometer must allow for mounting on a tripod using an M8 or M6 or M4 screw. - The total weight of the spectroradiometer must be less than 1.2 kg. - Remote control from PC via USB - Control software must also be supplied in the price of the Spectroradiometer - The spectroradiometer must be delivered calibrated, Calibration uncertainty must be better than Uc= 5,5% (k=2) for wawelenth 1400 ± 10 nm - Dimensions must be less than 150x150x150mm